Needle Microprobes
Fast-response needle probes for instant readings in tissue, semi-solids,liquids. Also for very small specimens, powders and materials
Product Enquiry
If you would like to send us an enquiry about this product, please click the button below, fill in the form and submit.
Product EnquiryPlease note the following definitions in the product descriptions below:
Time Constant is defined as the time required to reach 63% of final temperature in liquid. An accurate reading is obtained when the reading stops changing, typically 5-10 time constants.
Maximum Temperature refers to intermittent use of probe. When probe is used for continuous measurements, maximum temperature should be reduced by 25% unless specified.
Needle Microprobes
Fast-response needle probes for instant readings in tissue, semisolids and liquids. Also suitable for very small specimens, powders and materials.
The needle tip is sealed to ensure only stainless steel contacts the specimen. Sensors are accurate to ±0.1°C in the physiological range and have a maximum temperature of 200°C. Each sensor is fitted with a 5 ft. lead.
Probes are available in a range of needle gauges and lengths. The smallest microprobes (narrow gauge) give fastest reading. Short probes are easier to insert and last longer. Type #’s indicate needle gauge and needle length in cm (eg MT-29/1HT is a 29 gauge needle of length 1cm). An HT extension indicates the autoclavable version.
Sizes Available (ga/cm.):
Gauge | Sensors | Time constant |
29 ga (0.013”) | MT-29/1, MT-29/2, MT-29/3, MT-29/5 | 0.025s |
26ga (0.018”) | MT-26/2, MT-26/4, MT-26/6 | 0.1s |
23ga (0.025” | MT-23/3, MT-23/5, MT-23/8 | 0.15s |
All needle probes are non-isolated. Isolated versions of 23 ga. and larger needles are available for special order.
MT-4 Needle Microprobe
Similar to MT-29/1 but has blunt tip. Good for instant skin and surface temperatures as well as liquids. The MT-4 has a time constant of 0.025 secs and is not isolated.
Sensor Tip Diameter | 29 ga. (.013”) Blunt tip |
Accuracy | ±0.1°C in the physiological range |
Time Constant: | 0.025 seconds |
Maximum Temperature: | 200°C |
Minimum Temperature | -50°C |
Lead Length | 5 feet |
Sterilization Protocol | Gas ETO, Gluteraldehyde, Isoprophyl Alcohol(can be custom made to be autoclaveable) |
MT-D Needle Microprobe
Fast-responding, sealed blunt tip surface probe. Ideal for dental application for locating inflammation.
Sensor Tip Diameter | 29 ga. (.013”) |
Accuracy | ±0.1°C in the physiological range |
Time Constant: | 0.025 seconds |
Maximum Temperature: | 200°C |
Minimum Temperature | -50°C |
Lead Length | 5 feet |
Sterilization Protocol | Gas ETO, Gluteraldehyde, Isoprophyl Alcohol(can be custom made to be autoclaveable) |
ICT-4 Micro-Surface Microprobe
Micro-Surface Microprobe ideal for temperature measurements of integrated circuits and other micro specimens with 0.1°C accuracy.
0.5” straight stainless steel shaft with slightly protruding sensor bead to facilitate temperature probing. 5” tubular handle fits micro-manipulators. Maximum temperature 200°C (intermittent) or 100°C for continuous use.
Sensor Tip Diameter | .013” |
Accuracy | ±0.1°C in the physiological range |
Time Constant: | 0.025 seconds |
Maximum Temperature: | 200°C |
Minimum Temperature | -50°C |
Lead Length | 5 feet |
Sterilization Protocol | Gas ETO, Gluteraldehyde, Isoprophyl Alcohol |